Figure 3 shows the isolated probe response obtained by subtracting the flicker alone response from the combined flicker-plus-probe response. The left panel (A, B, C) of
Figure 3 shows these responses after shifting them in time to align the probe trigger times at zero. This highlights the relative change in the response amplitude to the probe at different times during the flicker cycle. The right panel of
Figure 3 shows the same responses on a larger time scale.
According to the theory of the paired flash technique, the isolated probe response can be used to estimate the amount of circulating current shut down by the test (flicker) stimulus at different times during the flicker cycle. To demonstrate the difference in the isolated probe response derived at 0 ms and at 33, 16, and 8 ms (for 15, 30, and 60 Hz, respectively), these responses have been plotted as a dashed line. The larger amplitude response at the start of the cycle indicates that the flicker stimulus minimally reduced the circulating current at this time, leaving most of the remaining current to be shut down by the probe flash. Similarly, near the middle of the cycle, the isolated probe response was relatively small since, at this time, a larger amount of circulating current had been shut down by the flicker stimulus, with less current left to be shut down by the probe flash.